12.9 Temperature of medium

Important
Please note the information in the section titled "Explosion
protection".
Compliance with the permissible temperature range for the
gaskets is mandatory.
FV4000-VT4/VR4
Standard
-55 ... 400 °C
HT design
(-67 ... 752 °F)
12.10 Flowmeter insulation
The pipeline may be insulated up to a maximum of 100 mm (4 inch)
upper edge.
Use of trace heating
Trace heating may be used under the following conditions:
If it is fixed directly on or around the pipeline
If, in the case of existing pipeline insulation, it is installed inside
the insulation (the maximum height of 100 mm (4 inch) must not
be exceeded)
If the maximum temperature the trace heating is able to produce
≤ the maximum temperature of the medium
The requirements to be met by integrators set out in EN 60079-
14 must be complied with!
Please note that the use of trace heaters will not impair EMC
protection or generate additional vibrations.
Fig. 59:
Flowmeter insulation
1
Maximum 100 mm (4 inch)
Change from one to two columns
D184B097U02
FS4000-ST4/SR4
-55 ... 280 °C (-67 ... 536 °F)
-
12.11 Ambient conditions
Resistance to climate to DIN 40040
Permissible ambient temperature range
Explosion protection / Model
None /
VT40 and VR40 / ST40 and
SR40
Ex ib /
VT41 and VR41 / ST41 and
SR41
Ex ia /
VT4A and VR4A / ST4A and
SR4A
Ex d /
VT42 and VR42 / ST42 and
SR42
FM
/
C
US
VT43 and VR43 / ST43 and
SR43
1) Category 2D (dust-ignition proof) maximum 60° C (140° F)
Permissible air humidity
Design
Standard
Climate-proof
70
1
60
50
40
30
20
10
0
-10
-20 opt. -55
-50
0
Fig. 60:
Relationship between the temperature of the fluid and the
ambient temperature
1
Ambient temperature
2
Media temperature
3
Permissible temperature
range for standard design
(≤ 280 °C (≤ 536 °F))
1
)
For the supply circuit (terminals 31 / 32) and the switching outputs 41 and 42,
cables suitable for temperatures up to T = 110 °C (230 °F) may be used without
restriction. Cables which are only suitable for temperatures up to
T = 80 °C (176 °F) restrict the temperature ranges. These restrictions also
apply to the VR version (remote design) and the PROFIBUS PA design with
plug connector.
Important
The legibility of the display can be impaired at temperatures
< 0 °C (< 32 °F) and > 55 °C (> 131 °F). The functionality of
the meter and the outputs remains unaffected by this. Please
refer to the order information for ambient temperatures < -
20 °C (< -4 °F).
Please note the information in the section titled 13
„Transmitter specifications".
FV4000 / FS4000
Temperature range
-20 ... 70 °C (-4 ... 158 °F)
-55 ... 70 °C (-67 ... 158 °F)
-20 ... 70 °C (-4 ... 158 °F)
-40 ... 70 °C (-67 ... 158 °F)
-20 ... 60 °C (-4 ... 140 °F)
-30 ... 60 °C (-40 ... 140 °F)
-20 ... 60 °C (-4 ... 140 °F)
-40 ... 60 °C (-40 ... 140 °F)
-20 ... 70 °C (-4 ... 158 °F)
-45 ... 70 °C (-49 ... 158 °F)
Humidity
Relative humidity max. 85 %,
annual mean ≤ 65 %
Relative humidity
≤ 100 % permanent
3
4
4
4
50
100
150 160
200
250
4
Installation for medium
temperature > 150 °C
(302 °F)
5
HT design
(≤ 400 °C (≤ 752 °F)),
FV4000-VT4 only
Specifications
1)
1)
1)
5
280
400
2
111