Approval Data For Hazardous Areas - ABB FV4000 Operating Instruction

Vortex/swirl flowmeter
Table of Contents

Advertisement

Ex relevant specifications

11.2.2 Approval data for hazardous areas

Power supply
Terminals 31, 32
circuit
Zone 1: Ex ib IIC
T
= (-40 °C) -20 ... 70 °C
amb
U
= 28 V
i
I
= 110 mA
i
P
= 770 mW
i
Effective internal capacitance: 14.6 nF
Type of protection
Effective internal capacitance
U
= 60 V
to ground:
m
Effective internal inductance: 0.27 mH
Zone 2: Ex nA [nL] IIC
T
= (-40 °C) -20 ... 70 °C
amb
U
= 14 ... 46 V
B
Zone 21 / 22: Ex tD A21 / Ex tD A22
T
= -20 °C ... 60 °C
amb
Power supply
Terminals 41, 42
circuit
Zone 1: Ex ib IIC
U
= 15 V
i
I
= 30 mA
i
P
= 115 mW
i
Effective internal capacitance: 11 nF
Effective internal capacitance
Type of protection
to ground:
U
= 60 V
m
Effective internal inductance: 0.14 mH
Zone 2: Ex nA [nL] IIC
U
= 16 ... 30 V
B
I
= 2 ... 15 mA
B
Zone 21 / 22: Ex tD A21 / Ex tD A22
T
= -20 °C ... 60 °C
amb
The devices must be installed in a protected environment in
accordance with the specific conditions on the test certificate.
Pollution degree 3 (see IEC 60664-1) must not be exceeded for the
macro environment of the device. The devices conform to degree of
protection IP 65 / IP 67. If the device is installed as intended, this
requirement is met by the housing as standard.
When connected to the line supply / not connected to the line supply,
the electrical circuits must not exceed overvoltage category III / II.
Change from one to two columns
100
11.2.3 Medium temperatures / Temperature
For the supply circuit "Terminals 31, 32" and the switching outputs
"Terminals 41, 42", cables suitable for temperatures up to T = 110 °C
(T = 230 °F) can be used without restriction.
Category 2/3G
For cables suitable only for temperatures up to T= 80 °C (T = 176 °F),
the interconnection of both circuits needs to be taken into account in
the event of a fault. Otherwise, the restricted temperature ranges
listed in the following table shall apply.
24.4 nF
Category 2D
For cables suitable only for temperatures up to T= 80 °C (T = 176 °F),
the restricted temperature ranges listed in the following table shall
apply.
Ambient
temperature
(-40) -20 ... 70 °C
((-40) -4 ... 158 °F)
(-40) -20 ... 70 °C
((-40) -4 ... 158 °F)
(-40) -20 ... 60 °C
19.6 nF
((-40) -4 ... 140 °F)
(-40) -20 ... 55 °C
((-40) -4 ... 131 °F)
(-40) -20 ... 50 °C
((-40) -4 ... 122 °F)
(-40) -20 ... 40 °C
((-40) -4 ... 104 °F)
1)
Medium temperatures > 280 °C (> 536 °F) with FV4000 Vortex flowmeter only
2)
The permissible limits for the ambient temperature are approval- and order-specific
(standard: -20 °C (-4 °F)).
3)
Category 2D (dust-ignition proof) maximum 60° C (140° F)
Maximum medium
temperature
FV4000 / FS4000
classes
Max. temperature
2)
at used
connecting cable,
"Terminals 31,
32", "Terminals
41, 42"
3)
110 °C (230 °F)
3)
3)
3)
80 °C (176 °F)
Temperature class
130 °C (266 °F)
195 °C (383 °F)
290 °C (554 °F)
400 °C (752 °F)
Max. permissible
medium
temperature
280 °C / 400 °C
1)
(536 °F / 752 °F)
1)
160 °C (320 °F)
240 °C (464 °F)
280 °C (536 °F)
320 °C (608 °F)
1)
400 °C (752 °F)
1)
T4
T3
T2
T1
D184B097U02