Table 6-2. Configuration Error Details (D:)
Message
Lower Range Value or
Upper Range Value is out
of limits
Configured Measurement
Mode not Supported
Volume Configuration
Error
Unsupported Probe and
Nozzle Configuration
Parameter Out of Limits
Low Low Level Alert
configuration is invalid
Low Level Alert
configuration is invalid
High Level Alert
configuration is invalid
High High Level Alert
configuration is invalid
User Defined Alert
configuration is invalid
Signal Quality Alert
configuration is invalid
No user configuration
errors detected
Service and Troubleshooting
PRELIMINARY
Description
The lower/upper range value is outside the
lower/upper sensor limits. This is outside the
range where the sensor works properly, hence
the measurement may be unreliable.
The configured Measurement Mode does not
work since support has not been purchased.
The volume cannot be calculated correctly
with the current configuration.
Rigid Single Lead 0.5 in / 13 mm probe is
configured in combination with 1.5 inch
Nozzle. This combination is invalid and results
in unreliable measurement.
One or more of the configuration values, in a
set that belongs together, has been changed
and reduced the measurement range of the
device.
Low Low Level Alerts will not be raised as
expected because of invalid configuration.
Low Level Alerts will not be raised as expected
because of invalid configuration.
High Level Alerts will not be raised as expected
because of invalid configuration.
High High Level Alerts will not be raised as
expected because of invalid configuration.
User Defined Alerts will not be raised as
expected because of invalid configuration.
Signal Quality Alerts will not be raised as
expected because of invalid configuration.
No user configuration errors detected.
Service and Troubleshooting
Recommended actions
1. Check the lower and upper range values in relation
to the sensor limits.
1. Upgrade the device.
2. Change the Measurement Mode to match, refer to
"Measurement mode" on page
1. Check that the level-volume values in the strapping
table are entered in increasing order, refer to
"Strapping table" on page
171.
2. Check that the number of strapping points to use is
correct.
3. Check size measures in the Volume Setup.
1. Install a nozzle with a supported inner diameter and
re-configure transmitter.
1. Check the configured values for Probe Length,
Vapor Dielectric Constant, and Upper Product
Dielectric Constant.
2. Restore default settings and reconfigure the device.
1. Check entered limit and deadband values in
comparison with entered tank and probe
parameters such as probe length, tank height etc.
1. Check entered limit and deadband values in
comparison with entered tank and probe
parameters such as probe length, tank height etc.
1. Check entered limit and deadband values in
comparison with entered tank and probe
parameters such as probe length, tank height etc.
1. Check entered limit and deadband values in
comparison with entered tank and probe
parameters such as probe length, tank height etc.
1. Check entered limit and deadband values in relation
to the selected variable.
1. Check entered limit and deadband values regarding
their range [0.1 – 10.0] and mutual relationship.