Emerson Rosemount 3308A Reference Manual page 193

3308 series wireless guided wave radar
Hide thumbs Also See for Rosemount 3308A:
Table of Contents

Advertisement

Reference Manual
00809-0100-4308, Rev CA
Table E-4. Advisory Alerts (A:)
Message
Database Memory Warning
Non-Critical User Data Warning
Volume Range Warning
Verification Mode Active
Button Stuck
HiHi Level Alert
Hi Level Alert
Lo Level Alert
LoLo Level Alert
User Defined Alert
1.
Location of the Alert in the HART command 48 Additional Status field.
Alert Message Mapping
PRELIMINARY
Additional status
Byte 0::Bit 2
Byte 2::Bit 1
Byte 4::Bit 7
Byte 4::Bit 4
Byte 1::Bit 5
Byte 5::Bit 4
Byte 5::Bit 5
Byte 5::Bit 6
Byte 5::Bit 7
Byte 5::Bit 3
Alert Message Mapping
(1)
Description
The device has failed to write to the
database memory at some time in the
past. Any data written during this time
may have been lost.
A user written parameter does not match
expected value.
The level measurement is outside the
configured volume range.
The device is in verification mode and is
not reporting actual information.
The button on the Electronics Board is
detected as stuck in the active position.
The level is above the defined limit.
The level is above the defined limit.
The level is below the defined limit.
The level is below the defined limit.
The variable has surpassed the user
defined limit.
August 2017
181

Advertisement

Table of Contents
loading

Table of Contents