Table 6-4. Advisory Alerts (A:)
Message
Database Memory
Warning
Non-Critical User Data
Warning
Volume Range Warning
Verification Mode Active
Button Stuck
HiHi Level Alert
Hi Level Alert
Lo Level Alert
LoLo Level Alert
User Defined Alert
Simulation Active
Service and Troubleshooting
PRELIMINARY
Description
The device has failed to write to the database
memory at some time in the past. Any data
written during this time may have been lost.
A user written parameter does not match
expected value.
The level measurement is outside the
configured volume range.
The device is in verification mode and is not
reporting actual information.
The button on the Electronics Board is
detected as stuck in the active position.
The level is above the defined limit.
The level is above the defined limit.
The level is below the defined limit.
The level is below the defined limit.
The variable has surpassed the user defined
limit.
The device is in simulation mode and is not
reporting actual information.
Service and Troubleshooting
Recommended actions
1. If logging dynamic data is not needed, this advisory
alert can be safely ignored.
2. Restart the device.
3. Reconfirm all configuration items in the device.
4. Restore default settings and reconfigure the device.
5. If the condition persists, replace the device.
1. Restart the device.
2. Reconfirm all configuration items in the device.
3. Restore default settings and reconfigure the device.
4. If the condition persists, replace the device.
1. Check volume configuration.
1. If this behavior is not desired, stop verification
mode.
1. Check the buttons for obstructions.
2. If the condition persists, restart the device.
3. If the condition persists, replace the device.
1. Bring the system to a safe state.
2. Verify that the level is within specified limits.
3. Reconfirm the level alert limit.
4. If not needed, disable this alert.
1. Bring the system to a safe state.
2. Verify that the level is within specified limits.
3. Reconfirm the level alert limit.
4. If not needed, disable this alert.
1. Bring the system to a safe state.
2. Verify that the level is within specified limits.
3. Reconfirm the level alert limit.
4. If not needed, disable this alert.
1. Bring the system to a safe state.
2. Verify that the level is within specified limits.
3. Reconfirm the level alert limit.
4. If not needed, disable this alert.
1. Bring the system to a safe state.
2. Verify that the process variable is within user
specified limits.
3. Reconfirm the user defined alarm limit.
4. If not needed, disable this alert.
1. If this behavior is not desired, stop simulation
mode.
2. If the condition persists, restart level
measurements.
August 2017