Configuration Parameters
August 2017
Figure D-5. Probe Angle ( )
Remote housing
If the transmitter head is mounted apart from the probe, the length of cable between probe and remote
housing must be configured.
User defined probe settings
Parameters for user defined probe.
Note
These settings should only be modified for customized probes. The settings are typically provided by
factory.
Geometry
Calibration offset
Difference between surface distance measured by device and the same distance measured by e.g.
handgauging with a measurement tape. A positive Calibration Offset value will increase the presented
Level value.
Show level below probe end as zero
When this setting is selected and the product surface is at or below the probe end, the level
measurement output will be zero.
Note
Only applicable for negative probe end peak.