Battery Recharge Intervals; Calibrate The Battery; Diagnostics And Troubleshooting; Factory Settings - Beckman Coulter MET ONE 3411 User Manual

Table of Contents

Advertisement

Table 5 Battery LED color indications
LED state
LED color
Flashing
Orange
Flashing
Green
Solid
Green

Battery recharge intervals

Table 6
shows the charge frequency that will increase battery life and
increase the interval between battery calibrations.
Table 6 Suggested battery recharge interval
Frequency of
use
0.5
Daily
Charge
weekly
Weekly
Charge
monthly
Monthly
Store on
charge
When the particle counter is not is use, batteries will slowly discharge
because of background processes on the instrument.

Calibrate the battery

The lithium ion Smart Battery will tolerate frequent partial discharges.
After many partial discharges, the accuracy of the battery gauge is
decreased.
To calibrate the battery charge gauge, set the Smart Charger to
calibration mode during discharge.
40 English
Battery status
Charge status
Low power
Not charging
Low power
Charging
Charged
Charging
Hours of sampling
1
2
Charge
Charge daily
Charge daily
weekly
Charge
Charge
Charge
weekly
weekly
weekly
Store on
Store on
Store on
charge
charge
charge

Diagnostics and Troubleshooting

The Diagnostics screen shows information that may be needed
for troubleshooting.
the Diagnostics screen.
Table 7 System Diagnostics screen example - Clock battery failure
Signal
Calibration
Flow
Clock battery
Battery 1 (bottom)
6
Battery 2 (top)
Laser current
For troubleshooting that involves technical support from the
manufacturer, the user may need to send a fax of the system status
printout to technical support.
1. On the Counter Navigation screen, push DIAGNOSTICS. Information
about the system, such as serial number and software version,
shows.
2. On the Diagnostics screen, push
The printout will show the serial number, date and time and other data
about the system.

Factory settings

For troubleshooting that involves technical support from the
manufacturer, the user may need to send a copy of the factory settings
to technical support.
Note: Factory settings cannot be changed by the user.
Table 7
shows an example of a failure notification on
Value
0.00 VDC
0.00 VDC
0.00 VDC
16.42 VDC
16.44 VDC
N/A
.
Status
OFF
OFF
FAIL
PASS
PASS
– –

Advertisement

Table of Contents
loading

Table of Contents