Set Operator Id And Password - Accumetrics VerifyNow User Manual

Blood microcoagulation system
Table of Contents

Advertisement

8.6

Set Operator ID and Password

The VerifyNow Instrument has been designed with features that restrict instrument
access. If this feature is enabled, a user identification number and a matching
password must be entered before an operator can use the instrument. In addition, it
is also possible to assign different authority levels to each user, depending upon the
instrument functions delegated to each user (e.g. running patient samples, QC
procedures or changing instrument settings).
The instrument's default setting does not provide for operator identification; however,
use the following procedure to adjust the setting or add/remove users based on your
institution's policy:
From the main menu, enter
1.
the keypad to enter the Operator ID number at the prompt. Press the Tab key to
advance to the Password prompt to enter the password number.
Press the Maintenance
2.
navigating within the Maintenance section.
Press the Password
3.
8.6.1 for a description of features of the Operator Log.
To enable or disable operator identification/passwords, press either the (← ← ← ← )
4.
or (→ → → → ) key to change the setting between Yes and No.
To add a user, press the Tab key to advance to "Oper" prompt.
5.
Record an Operator ID by entering up to 5 numbers using the keypad. If
an error is made during entry, use the (→ → → → ) key to move the cursor to the
mistake, then press the (← ← ← ← ) key to erase the mistake. Up to 100 users
may be entered.
Next, press the Tab key to advance the cursor to the password setting
(Pswd). Enter a password (up to 5 characters) using the keypad.
Next, press the Tab key to advance the cursor to the authority setting. Use
either (←,→) to scroll through the list of authority levels. If you do not wish
to establish authority levels for all users then choose the blank option
when scrolling through the authority level lists.
Once all information is entered for an operator, press the Add Item
key to add the operator to the list.
NOTE: At least one operator must have P,M,R,Q,A authority to access the
Operator Authority Log. Record operator names, operator IDs, and password
information and store the information with the instrument files.
VerifyNow System User Manual
Operator ID and
key. Refer to the Appendix for information on
key to display the Operator Authority Log. See Section
Page 77
INSTRUMENT SETTINGS - 8
Password, if required. Use
14340.J

Hide quick links: