Accumetrics VerifyNow User Manual page 35

Blood microcoagulation system
Table of Contents

Advertisement

Figure 5-16 WQC Level 1 Result
Open the cover. Remove the assay device and tube together in one piece by
19.
grasping the assay device finger grip and pulling straight up (Figure 5-17). Never
separate the tube from the assay device. Close the cover. Discard the assay
device and tube in a biohazard waste in accordance with institutional guidelines.
Figure 5-17 Remove Device
At the conclusion of the test, the resulting value will be displayed on the screen.
20.
Look up the expected values printed on the assay device pouch for Control 1
(Figure 5-18).
Figure 5-18 Interpreting WQC Level 1 Results
Compare the WQC result to the Control Level 1 range, and verify the result falls
21.
within the expected range. If the result falls within the expected range, the
VerifyNow System User Manual
Page 35
QUALITY CONTROLS - 5
Control Level
1 Range
14340.J