Wet Quality Control Results Log - Accumetrics VerifyNow User Manual

Blood microcoagulation system
Table of Contents

Advertisement

Feature
Description
The failed parameter(s) will be indicated by an arrow (←) to the
Flagged
Parameter
right of the measured value(s). Refer to the Appendix for a
description of parameters.
In addition to viewing earlier results, icons keys enable results to be printed
(optional).
Icon
Print Result
Print All
Results
Back
7.2.2

Wet Quality Control Results Log

The cursor is positioned on the most recent assay result (Figure 7-5). Navigate up
and down an individual list using the (← ← ← ← , → → → → ) keys.
VerifyNow System User Manual
O1:
High optical signal level
O2:
Low optical signal level
V1:
Maximum vacuum level
V2:
Maximum leak rate
P1:
Maximum pressure level
F1:
Mixer frequency flag count
F2:
Mixer frequency transitions high-to-low
Mask: Identifies failing component element
Description
Scroll to the selected record, and select this icon to print the
selected EQC results.
Scroll backwards through the results using the
select this icon to print all EQC results in the log that are
forward of the cursor position.
Select this icon to return to the previous screen.
Page 64
RETRIEVING TEST RESULTS - 7
← ← ← ←
key, and
14340.J

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents