Retrieving Test Results and Performance
7
Data
The instrument stores information on patient test results, quality control tests,
troubleshooting events, and usage statistics. This section covers the activities to
retrieve stored information and display it on the screen, print it, or export it to a
network (where network connectivity is enabled). Table 7-1 lists and describes the
six activities available to an operator.
Table 7-1 Retrieving Test Results and Performance Data
Activity
Retrieve patient
results
View WQC results
(Level 1 & 2)
View EQC
diagnostic results
View usage log
View alarms and
errors log
Transfer data
across a network
7.1

Retrieving Patient Test Results

The instrument stores the last 150 patient results in its memory. You can retrieve and
print information about the patient result, assay type, assay date and time, and
operator identification (when configured).
Patient information can be retrieved by date and time or by patient identification,
depending on the instrument setting. Refer to Section 7.1.2 if the instrument is
configured to use patient identification; otherwise, refer to 7.1.1 for the appropriate
procedure.
7.1.1

Patient Information Stored by Date/Time

From the main menu, enter
keypad to enter the Operator ID number at the prompt. Press the Tab key to advance
VerifyNow System User Manual
This feature enables you to retrieve the last 150 patient results from
earlier tests. Refer to Section 7.1.
Use this feature to view earlier WQC results. This feature enables
you to retrieve the last 100 WQC results. Refer to Section 7.2.
Use this feature when troubleshooting instrument performance.
This feature enables you to retrieve the last 100 EQC results. Refer
to Section 7.2.
This feature enables you to view the quantity and type of assays
performed on the instrument. Refer to Section 7.3.
Use this feature when troubleshooting instrument performance.
This feature enables you to retrieve information on recent alarms
and errors. Refer to Section 7.4.
This optional accessory enables you to transfer information across
a network. This includes all assay, WQC and EQC results as well
as messages, alerts, alarms, and errors logged. Refer to Section
7.5.
Operator ID and
Page 59
RETRIEVING TEST RESULTS - 7
Description
Password, if required. Use the