Electronic Quality Control - Accumetrics VerifyNow User Manual

Blood microcoagulation system
Table of Contents

Advertisement

Level 2
Run before the first use of each new lot
WQC
(or shipment) of assay device kits.
Run during troubleshooting to resolve
unexpected events. Refer to Chapter 10,
Troubleshooting.
Run when the assay device kit
temperature indicator shows exposure to
elevated temperatures. Refer to Chapter
4, Reagent Storage and Handling.
Run every 30 days (only for IIb/IIIa
assay).
Use as part of your institution's quality
control program
The following sections provide a detailed description of the procedures.

Electronic Quality Control

5.1
Electronic Quality Control (EQC) is the primary quality control mechanism for the
VerifyNow Instrument. It consists of a re-usable device that is inserted by the
operator into the assay port and is used to perform a comprehensive testing routine
that confirms performance of key instrument components.
This procedure involves a reusable EQC device and takes up to two minutes. If a
fault is detected in any of its systems, the instrument cannot perform patient testing
until the fault is corrected.
Use the following procedure to perform an EQC:
NOTE: Pressing the Back
the main menu.
Power on the instrument using the power switch on the back panel ("I" designates
1.
on). The instrument will power on and perform a self-testing routine lasting
approximately 30 seconds. After the self-testing is complete, the start screen will
display. Press the Next
source not found.). The instrument should be allowed to warm-up for at least 15
minutes prior to use.
VerifyNow System User Manual
key will stop the EQC test at any point and return to
key to advance to the main menu (Error! Reference
Page 26
QUALITY CONTROLS - 5
Verifies instrument performance by
checking that actual results are within
a higher range of values.
14340.J