Accumetrics VerifyNow User Manual page 103

Blood microcoagulation system
Table of Contents

Advertisement

Parameter
Issue Description
P1
Maximum pressure
level
F1
Mixer frequency flag
count
F2
Mixer frequency
transitions high to low
MASK
Failing optical element
VerifyNow System User Manual
Resolution
This event is logged as Error 19 in the event log.
1) Use the cleaning device (see Section Error! Reference
source not found.) then perform an EQC (see Section 5.1).
2) Repeat if necessary.
3) Call Customer Support if EQC does not pass after three
attempts.
1) Check the alarm and error log (see Section 7.4.1) and note
the frequency of Error 16 in the log.
2) Call Customer Support with this information.
1) Visually inspect the EQC device to verify it has not been
damaged. If you have another EQC device, repeat EQC using
the second device.
2) Check the alarm and error log (see Section 7.4.1) and note
the frequency of Error 16 in the log.
3) Call Customer Support with this information.
This event is logged as an error in the event log (e.g. #4, #29).
It commonly occurs when the EQC device has not been
removed completely from the assay device port when
prompted or when the instrument is located in adverse lighting
conditions (refer to installation guidelines).
Call Customer Support if the EQC device had been removed
completely at the end of EQC and it fails.
Page 103
TROUBLESHOOTING - 10
14340.J