At the conclusion of the test, the resulting value will be displayed on the screen.
18.
Look up the expected values printed on the assay device pouch for Control Level
2 (Figure 5-32).
Figure 5-32 Interpreting WQC Level 2 Result
Compare the WQC result to the Control Level 2 range, and verify the result falls
19.
within the expected range. If the result falls within the expected range, the
instrument has passed WQC Level 2. If the WQC result does not fall within the
stated range, perform an EQC test to ensure that the instrument is working
properly. If the EQC is OK, prepare a new WQC sample and repeat the WQC
procedure with a new assay device. If the EQC is not OK, follow the instructions
in Chapter 10, Troubleshooting. If the WQC result fails on the second attempt,
contact Accumetrics' Customer Support.
Press the Next
20.
VerifyNow System User Manual
key to return to the main menu.