Figure 5-29 Assay Progress
CAUTION: The sample is pressurized during the assay. Do not
remove the sample tube or assay device during the assay. To abort,
press the Back
assay device along with the tube.
When the results display, record the results or print
16.
instrument has an attached printer (Figure 5-30).
Figure 5-30 WQC Level 2 Result
Open the cover. Remove the assay device and tube together in one piece by
17.
grasping the assay device finger grip and pulling straight up (Figure 5-31). Do not
separate the tube from the assay device. Close the cover. Discard the assay
device and tube in a biohazard waste in accordance with institutional guidelines.
Figure 5-31 Remove Device
VerifyNow System User Manual
key. Wait for the prompt before removing the
Page 41
QUALITY CONTROLS - 5
the results, if your
14340.J