Enable Patient Id - Accumetrics VerifyNow User Manual

Blood microcoagulation system
Table of Contents

Advertisement

Figure 8-4 Time Out Setting
Time Out Setting
Press the Tab key 3 times to advance the cursor to the Time Out setting. The
4.
instrument supports three different time out settings, depending on the operation
being performed. These settings correspond to the following three different fields
(Table 8-2).
Table 8-2 Time Out Settings
Field
Activities Affected
1
Affects all input operations associated with an
assay or QC.
2
Affects the time that a user's password will
remain on the screen (when configured).
3
When idle, affects when the screen saver
"stand-by" feature is engaged. When engaged,
the "stand-by" feature returns to the main
menu and dims the display.
Record the new timeout settings for the three fields – between 0 and 255
5.
minutes. Assigning a timeout value to 0 means the affected screens will not time
out. Press the Tab key to move between the settings.
NOTE: Unless the third field is set to 0, the second field needs to be less than or
equal to the third field.
Press the Save
6.
menu. Press the Back
the changes.

Enable Patient ID

8.5
The VerifyNow Instrument supports patient result traceability. This includes a feature
that assigns a patient identification (ID) to each patient sample. The instrument's
default setting does not assign patient ID; however, use the following procedure to
adjust the setting:
VerifyNow System User Manual
key to keep the new setting and return to the Maintenance
key twice to return to the Main Menu without saving
Page 75
INSTRUMENT SETTINGS - 8
Save Setting
Back
Default Time Setting
2 minutes
3 minutes
5 minutes
14340.J