Reagent Storage And Handling; Storage And Stability - Accumetrics VerifyNow User Manual

Blood microcoagulation system
Table of Contents

Advertisement

Reagent Storage and Handling

4
This section describes special considerations for receiving, storing and handling of
VerifyNow reagents and devices.

Storage and Stability

4.1
Store assay devices (in their unopened pouches) and quality control materials until
use at temperatures described in Table 4-1. Do not freeze.
Table 4-1 Recommended Storage Conditions
Assay
VerifyNow Aspirin assay device
VerifyNow P2Y12 assay device
VerifyNow IIb/IIIa assay device
VerifyNow Assay WQC pellets and
diluents
Assay devices and quality control materials are suitable for use until the expiration
date printed on the label. Do not use assay devices or WQC materials beyond the
expiration date. Refer to assay kit boxes and pouch for designated expiration date
and recommended storage conditions.
VerifyNow System User Manual
REAGENT STORAGE AND HANDLING - 4
Storage Temperature
2-25ºC (36-77ºF)
2-25ºC (36-77ºF)
2-8ºC – extended storage
18-25ºC (64-77ºF) - may be stored for up to 8
weeks, but not beyond the expiration date. When
removed from refrigeration, write the discard date
on the pouches and/or kit box
15-30ºC (59-86ºF)