Retrieving Quality Control - Accumetrics VerifyNow User Manual

Blood microcoagulation system
Table of Contents

Advertisement

Feature
Description
%
Operator ID
Displays the operator ID of the user who performed the test (when
configured).
Date and Time
Displays the date and time by year, month, day, hour (24 hr clock),
Stamp
and minute using the following notation, "YYMMDDHHMM."
To calculate percent inhibition for IIb/IIIa assays from the listed results, locate a
matching set of pre-drug PAU and post-drug PAU values for a patient using the date
and time stamp. Calculate the % inhibition using the following formula:
% Inhibition = (pre-drug PAU – post-drug PAU) X 100
In addition to viewing earlier results, icons keys enable results to be printed (optional)
or another assay started.
Icon
New Assay
Print Result
Print All
Results
Back
7.2

Retrieving Quality Control

The instrument stores the last 100 EQC and WQC results in its memory. You can
retrieve and print information about the earlier EQC parameters, WQC results, assay
type, date and time, and operator identification (when configured). Use the following
procedure to view the quality control results.
VerifyNow System User Manual
P2Y12 assay (PRU result)
P2Y12 assay (Base result)
P2Y12 assay (Percent Inhibition result)
pre-drug PAU
Description
Select this icon to perform another assay on the same
patient (when Patient ID is enabled). The instrument will
display the Patient ID screen with the Patient ID completed.
Scroll to the selected record, and select this icon to print the
selected assay results.
Scroll backwards through results using the ← key, and
select this icon to print all assay results in the log that are
positioned forward of the cursor, or all of the assay results
for a particular patient (when Patient ID function is
activated).
Select this icon to return to the Main Menu.
Page 62
RETRIEVING TEST RESULTS - 7
14340.J