From the main menu, enter
3.
the keypad to enter the Operator ID number at the prompt. Press the Tab key to
advance to the Password prompt to enter the password number.
Press the QC
4.
Figure 5-8 Start WQC Level 1
Open the foil pouch and remove the assay device just before use. Hold it by the
5.
finger grip.
NOTE: Each assay device has a finger grip. Avoid handling other surfaces of the
assay device.
Inspect the humidity indicator. The humidity indicator (small square) should be
6.
pink on one side and white on the other. If it is pink on both sides, discard and
use a new assay device (
NOTE: Do not try to re-attach any assay device needles that may inadvertently
7.
removed. If this occurs, discard the device and use a new one
Figure 5-9
8.
).
8.
Remove the needle's protective sheath by pulling directly up on the sheath (
9.
NOTE: Do not try to re-attach any assay device needles that may inadvertently
10.
removed. If this occurs, discard the device and use a new one
Figure 5-9
11.
). Do not twist the sheath, as this may remove the needle.
11.
VerifyNow System User Manual
Operator ID and
key. Wait for the image of the device to display (Figure 5-8).
Page 31
QUALITY CONTROLS - 5
Password, if required. Use