If sufficient time has elapsed, the instrument will prompt you to perform an EQC
4.
(Figure 6-4). Refer to Chapter 5, Quality Controls for more information.
Figure 6-4 EQC Required
After a PASS has been achieved on EQC the instrument will be ready to perform
an assay.
When the initial screen displays, press the Assay
5.
required, the patient ID prompt will display. If not, the instrument will prompt you
to insert the assay device (go to step 3).
A series of question marks will signify the required length of the patient ID
established by your institution. Use the numerical pad to input a numerical
patient ID and then press the Next
Figure 6-5).
Figure 6-5 Patient ID
NOTE: A question mark is not a valid input. All question marks must be replaced
by valid patient ID numbers in order to proceed with performing the assay.
To enter a letter for patient ID, press the Text
Use the up
screens. At the desired screen, press the (← ← ← ← ) or (→ → → → ) to move to the
desired character.
VerifyNow System User Manual
key (
and down
arrows to scroll through the text option
Page 48
PATIENT TESTING - 6
key. If the patient ID is
key.