PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04
Resuscitation Module
1. To use the Suction feature, turn it on using the On/Off
switch.
NOTE:
The gas supply must be turned on.
2. To adjust suction from 0 kPa to -19.99 kPa,, use the
Suction control knob.
3. Prior to use, verify that the suction gauge reads -19.99 kPa
(-150 mm Hg) ± 0.53 kPa (4 mm Hg) when suction
control is set to maximum and the suction output is
occluded. The suction output may be occluded at the
suction jar.
4. Verify that the suction switch turns the suction completely
OFF and completely ON.
WARNING:
There are potential hazards associated with the delivery
of supplemental oxygen. If it is necessary to administer oxygen, notify the attending physician
immediately. Failure to do so could result in infant injury.
WARNING:
Always monitor the airway pressure, and set appropriate pressure relief during infant
resuscitation. Failure to do so could result in infant injury.
WARNING:
For prolonged ventilation, use a heat and moisture exchanger or heated humidification. Failure
to do so could result in infant injury.
5. On units with the AutoBreath™ Infant Resuscitator, perform the following:
a. To turn the timing circuit on and off, use the On/Off switch.
b. To adjust the Positive End Expiratory Pressure (PEEP) in the patient circuit, use the PEEP
Min/Max control.
c. Adjust the inspiratory and expiratory times using the Rate Control.
5 - 12