WARNING:
The internal adjustable airway pressure relief is not functional with a self-inflating bag because
the patient supply is connected to the fresh gas reservoir on the self-inflating bag rather than
directly to the patient airway. When using a self-inflating bag, always use a suitable external
airway pressure relief valve connected to the patient airway. To obtain maximum flow, turn the
adjustable airway pressure relief knob to the Max. position. Failure to do so could result in infant
injury.
WARNING:
Using a breathing circuit other than the AutoBreath™ Infant Resuscitator Breathing Circuit
(PN 81 000 19/ MU10850) may inadvertently restrict the gas flow such that the indication of the
gas flow and the adjustable airway pressure relief may be inaccurate. Infant injury could occur.
• Patient Breathing Circuits—Use the patient breathing circuit (PN 81 000 19/ MU10850) in
conjunction with the AutoBreath™ Infant Resuscitator. Additionally, a patient supply circuit for
manual bagging, or blow-by, (PN 81 001 26/ MU10899) may be used.