PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04
17. Mount the controller on the upper post.
18. Install the two #8-32 x ¾" screws that secure the
controller to the head support plate.
19. Remove and discard the two #10-32 x 3/8" Nylok
screws.
20. Secure the lower rear cover to the upper post using the
six #10-32 x ¾" white screws.
21. Install the two #10-32 x 2.6" phillips head screws through the
two #10 lockwashers, the rear handle, and the two spacers to
secure the rear handle to the lower rear cover.
22. Secure the upper rear cover to the upper post using the four
#10-32 x ¾" white screws.
23. Connect the line cord to the power connector on the rear of the
controller.
CAUTION:
To prevent removal of the controller chassis with the AC
power on, secure the line cord to the back cover. Failure
to do so could result in equipment damage.
24. Install the #8-32 x 3/8" screw and the cable clamp provided to secure the line cord to the lower rear
cover.
25. Install any accessories ordered by using the installation instructions provided with each accessory.
26. To ensure proper operation of the Resuscitaire Birthing Room Warmer, perform the "Operational
Checkout Procedure" on page 4-14.