PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04
Symptom
Blender alarm activates.
Heater does not turn ON.
Pre-Warm Mode is not activated
upon a system reset.
Heater is ON, but the indicators
are OFF.
Cannot obtain the desired suction
pressure.
Blender O
concentration setting
2
does not match the blender set-
ting.
Blender output mixture is outside
the specified range (20.8% to
100%).
Patient Gas Supply measured
flow is outside of the specified
± 3% of full scale or 10% of set-
ting.
Patient Gas Supply pressure
relief gauge exceeds 50 cmH
± 10 cmH
O (4.9 kPa ± 1 kPa)
2
7 - 2
Possible Cause
Oxygen and/or air are discon-
nected.
Oxygen or air pressures are low.
Both tanks may not be opened.
Pressure differential between the
oxygen and air supplies exceed
207 kPa (30 psi) ± 14 kPa (2 psi).
Heating element is open.
There is an internal malfunction. Refer the unit to service.
System is in Diagnostic Mode.
There is an internal malfunction. Replace the power and control
There is an internal malfunction. Replace the display P.C. board.
Suction jar rim gasket is worn.
Suction jar is loose or chipped.
Connecting hose is bent or tan-
gled.
Hose is leaking between the reg-
ulator and the bottle.
Pressure or suction gauge is not
measuring correctly.
Blender knob is loose.
Blender has failed.
Air inlet H
O filter is missing or
2
clogged.
Blender has failed.
Dirty or clogged air/O
Pressure relief valve has failed.
O
2
Corrective Action
Connect oxygen and/or air.
Adjust oxygen or air pressure to
correct setting.
Check cylinder gauge.
Adjust oxygen or air pressure to
correct setting.
Refer the unit to service.
Exit Diagnostic Mode by press-
ing the reset button.
P.C. board.
Replace the suction jar.
Tighten or replace the suction jar.
Replace the hose.
Replace the hose.
Refer the unit to service.
Refer the unit to service. Do not
attempt to tighten the knob as the
unit will need recalibration.
Replace the blender or refer the
unit to service.
Replace or install filter.
Replace blender or refer the unit
to service.
hose.
Replace the hose.
2
Replace the pressure relief valve
or refer the unit to service.