PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04
Name
Baby Temperature display
°C/°F key
>37°C indicator
>37°C key
Set Temperature display
Up and Down arrow keys
Heater Power display
Apgar Timer Stop/Start and Reset
keys
5 - 2
Digital display of the infant's temperature in °C or °F, whether in
Manual or Baby Mode. If the skin temperature probe is discon-
nected from the controller, the display is blank in Manual and
Baby Mode. When in Pre-Warm Mode for software versions 1.08
or greater, the display indicates "PrE" regardless of whether or not
the probe is connected.
Press to alternately select °C or °F for the Baby Temperature and
Set Temperature displays.
Lights to indicate that the Temperature Override Mode is selected.
Press to place the Set Temperature display in Temperature
Override Mode. This key is inactive until the set temperature has
been set to 37.0°C (98.6°F).
In Baby Mode, it displays the set temperature as selected by the
Up and Down arrow keys in °C or °F as selected by the °C/°F key.
In Pre-Warm and Manual Modes, the display is blank if the skin
probe is not connected.
Serves 2 functions to regulate temperature:
In Manual Mode, adjusts heat output.
Press the Up arrow key to raise the heater power from 0% to
100% in 10% increments. Press the Down arrow key to lower the
relative heater power from 100% to 0% in 10% increments.
In Baby Mode, adjusts chosen temperature setpoint.
Press the Up arrow key to raise the set temperature from 34.0°C
(93.2°F) to 37.0°C (98.6°F). Press once to raise the set tempera-
ture in 0.1° increments. Press and hold to raise the set temperature
rapidly. Press the Down arrow key to lower the set temperature
from 37.0°C (98.6°F) to 34.0°C (93.2°F). Press once to lower the
set temperature in 0.1° increments. Press and hold to lower the set
temperature rapidly.
In Temperature Override Mode, press the Up arrow key to raise
the set temperature from 37.0°C (98.6°F) to 38.0°C (100.4°F).
Press the Down arrow key to lower the set temperature from
38.0°C (100.4°F) to 37.0°C (98.6°F).
When illuminated, the Keypad Lock key locks the Up and Down
arrow keys.
Displays the relative heater power in 10% increments from 0% to
100%.
Press the Stop/Start key to start or stop the Apgar Timer. When
the timer is running, press the Reset key to reset the timer to 0 and
restart the Apgar count. When the timer is stopped, press to turn
the Apgar Timer off. You can reset the Apgar Timer anytime.