Blender Module (Optional); Resuscitation Module (Optional) - Dräger Resuscitaire WBR82 Instructions For Use Manual

Radiant warmer product
Hide thumbs Also See for Resuscitaire WBR82:
Table of Contents

Advertisement

PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04
When operating the controller in Manual Mode, you can adjust the heater power from 0% to full power in
10% increments. After 10 min of operation in Manual Mode, the Check Patient alarm occurs. To silence
the alarm for 10 min, verify the infant temperature, and press the Silence/Reset key. Failure to
acknowledge the Check Patient alarm by pressing the Silence/Reset key within the next 5 min causes
the heater to shut down, and an alarm occurs.
When operated in Baby Mode, the controller uses a skin probe to adjust the heater output automatically.
The probe is connected between the controller input and the patient. This feature allows the warmer
module to maintain a selected set temperature.
The APGAR Timer displays the elapsed time and sounds a notification tone at 1, 5, and 10 min.
The Keypad Lock key, when pressed, activates or deactivates the Up/Down arrows, the Select Mode
key, and the >37 °C key.
A procedural silence timer blocks out Low Baby Temp acoustic alarm signals for 5 min during routine
procedures.
WARNING:
Do not place the device into use if the system fail alarm indicator remains on after the controller
diagnostic test is completed. Refer it to qualified service personnel. Failure to do so could result
in patient injury or equipment damage.
WARNING:
Do not place the device into use if any indicator fails to light, or any segments of the digital
display fails to light, or any audible tone fails to sound. Refer it to qualified service personnel.
Failure to do so could result in patient injury or equipment damage.

Blender Module (Optional)

The blender module provides blended gas from 21% to 100% to the patient outlet on the resuscitation
module.

Resuscitation Module (Optional)

The resuscitation module contains pneumatic outlets necessary for infant resuscitation. Controls and
displays for the module are located above the bassinet on the warmer column.
Three variations of resuscitation modules exist: Resuscitation Module with Internal Pressure Manometer
and 15 mm Fitting, Resuscitation Module with External Pressure Manometer, and Resuscitation Module
with AutoBreath™. All Resuscitation Modules have an adjustable airway pressure relief.
NOTE:
When pressure in the hose that delivers gas to the patient falls below -0.4 kPa (-4 cmH
valve installed at the Patient Outlet connection opens. In the unlikely failure of the gas supply, the one-
way valve allows patient to inspire room air.
The resuscitation modules consist of the factory-installed components discussed on the following pages.
2 - 2
O), the one-way
2

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Resuscitaire rw82Resuscitaire wmrw82

Table of Contents