System Fail Error Codes - Dräger WBR82 Instructions For Use Manual

Radiant warmer products
Hide thumbs Also See for WBR82:
Table of Contents

Advertisement

PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04
Symptom
AutoBreath™ Infant Resuscita-
tor measured BPM is outside of
the specified range.
The Inspiratory/Expiratory (I/E)
Ratio is Greater Than 1:2.4 and
Less Than 1.6:1, 1:1.

System Fail Error Codes

Error Number
Error #2
Error #3
Error #4
Error #5
Error #6
Error #7
Error #8
Error #9
Error #10
Error #11
Error #12
Error #13
Error #14
Error #15
Error #16
Error #17
Error #20
Error #21
There is an internal malfunction. Refer the unit to service.
Resuscitation Module has failed. Replace the resuscitation mod-
System Fail Error Codes
Explanation
AC line voltage too high
Solid State Relay's path shorted Replace the power and control P.C. board.
Heater circuit failure—short
Solid State Relay #1 failure—
open
Solid State Relay #2 failure—
open
Safety relay failure—short
AC heater current high
Hardware detects a high tem-
perature >39.5°C (103.1°F)
Power Fail circuit failure
CPU voltage out of specifica-
tion
Keypad failure
ROM test failure
RAM test failure
EEPROM test failure
Analog/Digital (A/D) offset out
of specification
Analog/Digital (A/D) gain out
of specification
External watchdog timer fail
Ambient temperature probe
error >0.4°C
Possible Cause
Ensure that the power source is in compliance with
the electrical specification of the system.
Replace the power and control P.C. board.
Replace the power and control P.C. board.
Replace the power and control P.C. board.
Replace the power and control P.C. board.
Inspect the quartz tube heating element. If neces-
sary, replace it. Otherwise, replace the power and
control P.C. board.
If the heater is on, replace the power and control
P.C. board. Otherwise, check for an ancillary heat
source.
NOTE:
This error is not included in the error log.
Replace the power and control P.C. board.
Check the voltage from the DC power supply.
Replace the display P.C. board.
Replace the power and control P.C. board.
Replace the display P.C. board.
Replace the power and control P.C. board.
Replace the display P.C. board.
Replace the display P.C. board.
Replace the power and control P.C. board.
Check for damaged ambient temperature probes.
The ambient temperature probe is internal, and must
be attached in order for the unit to function properly.
Corrective Action
ule or refer the unit to service.
Corrective Action
7 - 3

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Rw82Wmrw82