Resuscitation Module with AutoBreath™ Infant
Resuscitator (Optional)
WARNING:
Always monitor the airway pressure, and set appropriate
pressure relief during infant resuscitation. Failure to do
so could result in infant injury.
Automatic Patient Circuit with 15 mm Outlet
For automatic resuscitation, use the Automatic Patient Circuit,
(10 mm tubing with exhalation valve and exhalation valve
control line tubing).
WARNING:
Using a breathing circuit other than the one supplied for
the Patient Outlet or one that has tubing with a inner
diameter smaller than 10 mm may inadvertently restrict
the gas flow such that the indication of the gas flow and
the adjustable airway pressure relief may be inaccurate.
Infant injury could occur.
WARNING:
Potential hazards are associated with the delivery of supplemental oxygen. If it is necessary to
administer oxygen, notify the attending physician immediately. Failure to do so could result in
infant injury.
1. Prior to use, ensure the gas supply is turned off at the source.
2. Prior to patient use, turn the AutoBreath™ Infant Resuscitator circuit OFF using the On/Off switch.
3. Connect the Automatic Patient Circuit to the Patient Outlet connector.
4. Connect the exhalation valve control line tubing to the Exh Valve connector.
5. Turn the AutoBreath™ Infant Resuscitator circuit ON using the On/Off switch.
6. Using the Patient Supply Flow Rate control, adjust to the desired fresh gas flow.
7. To check the fixed internal Airway Pressure Relief control, set the desired airway pressure limit, and
block the exhalation valve port exhaust and the patient port of the exhalation valve.
Airway Pressure
8. To check the pressure limit, observe the Airway Pressure gauge.
9. Turn ON the AutoBreath™ Infant Resuscitator circuit using its On/Off switch.
10. Adjust the Rate (bpm) control to 18 breaths per minute (bpm).
11. To set the Positive End Expiratory Pressure (PEEP) threshold, block the patient port of the patient
breathing circuit. Do not block the exhalation valve exhaust port.
12. Observe the PEEP indicated on the Airway Pressure gauge, and use the PEEP control to adjust the
desired PEEP.
4 - 24