PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04
10. Check the Probe alarm:
a. Disconnect the skin temperature probe from the skin temperature probe connector. The Baby
Temperature display goes off, the Probe indicator flashes, and the alarm sounds.
b. Connect the skin temperature probe to the skin temperature probe connector.
11. Check the Apgar Timer:
a. Press the Start/Stop key. The Apgar Timer display comes on and begins counting up from 0 s.
b. Press the Start/Stop key. The Apgar Timer display stops counting.
c. Press the Reset key. The Apgar Timer display goes off.
12. Check the examination light:
a. Press the Examination Light key. The examination light comes on.
b. Press the Examination Light key. The examination light goes off.
13. Check the Power Failure alarm:
NOTE:
The unit must be connected to the AC line for at least 8 min before the power failure circuitry becomes
active.
a. Turn off the circuit breaker on the rear panel.
b. Turn ON the power switch on the front panel.
c. The Power Fail indicator comes on, and the alarm sounds.
d. Turn OFF the power switch, and turn on the circuit breaker.
e. Disconnect the skin temperature probe from the skin temperature probe connector.

Mechanical Operation

1. Check the bassinet tilt control (Resuscitaire Birthing
Room Warmer shown):
a. Support the rear lower edge of the bassinet with your
palm, and pull up on the lever located at the bottom
rear of the bassinet.
b. Place the bassinet in the 5° and then the 10° tilt
position.
c. Return the bassinet to the level position.
4 - 16