Dräger WBR82 Instructions For Use Manual page 18

Radiant warmer products
Hide thumbs Also See for WBR82:
Table of Contents

Advertisement

PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04
WARNING:
The internal adjustable airway pressure relief is not functional with a self-inflating bag because
the patient supply is connected to the fresh gas reservoir on the self-inflating bag rather than
directly to the patient airway. When using a self-inflating bag, always use a suitable external
airway pressure relief valve connected to the patient airway. To obtain maximum flow, turn the
adjustable airway pressure relief knob to the Max. position. Failure to do so could result in infant
injury
WARNING:
Using a breathing circuit other than the one supplied may inadvertently restrict the gas flow
such that the indication of the gas flow and the adjustable airway pressure relief may be
inaccurate. Infant injury could occur.
• Patient breathing and supply circuits—Use the disposable patient supply circuit (PN 81 001 27/
MU10900) for manual bagging on the Resuscitation Module.