PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04
5. Check the pass-through drawer (Resuscitaire Birthing
Room Warmer shown):
a. Slide the pass-through drawer in and out on both
sides of the cart and bassinet.
b. Return the pass-through drawer to its center position.
6. Check the warmer module swivel operation:
WARNING:
Push on the sides of the warmer module. Do not
touch under the warmer module. Personal injury
could occur.
a. Push on the sides of the warmer module, and
rotate it 90° to the left or right of center. Do not
touch under the warmer module.
WARNING:
For effective heating, ensure that the warmer
module is centered above the patient. Failure to
do so could result in personal injury.
b. Return the warmer module to its center
position.
7. Check the operation of the accessory x-ray cassette tray:
a. Grasp the middle of a side panel, and pull the x-ray
cassette tray out from under the bassinet.
b. Return the x-tray cassette tray to under the bassinet.
4 - 18