PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04
a. Connect the line cord to the power connector on the rear of the bassinet/cart assembly.
b. Connect the 102 cm (40") power cord provided between the connector on the front of the
bassinet/cart assembly and the power connector on the controller.
c. Using the cable clamp and the #8-32 x 3/8" screw, secure the 102 cm (40") power cord to the
lower rear cover.
23. Install the two #10-32 x 2.6" phillips head screws through the two #10 lockwashers, the rear handle,
and the two spacers to secure the rear handle to the lower rear cover.
24. Install any accessories ordered by using the installation instructions provided with each accessory.
25. Insert the pins of the back end panel in the corner brackets of the bassinet.
26. Slide the side panels straight down into their slots on the bassinet.
27. Install the front end panel:
a. Slide the front end panel into the front of the bassinet assembly until its release tabs catch.
b. Raise the front end panel to its fully raised position.
28. To ensure proper operation of the Resuscitaire Radiant Warmer (RW82), perform the "Operational
Checkout Procedure" on page 4-14.