X-Ray Cassette Tray Accessory (Optional); Universal Cylinder Bracket (Optional) - Dräger WBR82 Instructions For Use Manual

Radiant warmer products
Hide thumbs Also See for WBR82:
Table of Contents

Advertisement

PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04

X-Ray Cassette Tray Accessory (Optional)

1. To install an x-ray cassette, perform the following:
a. Swing the warmer module to the right or left of center as required to position the x-ray machine.
b. Lift the left or right bassinet side panel up.
c. Slide the x-ray cassette tray out.
d. Place the x-ray cassette on the tray.
e. Return the x-ray cassette tray to the bassinet.
f. Align the x-ray cassette tray as desired with the markings on the x-ray cassette and the relative
markings on the inside of the bassinet walls.
2. When the x-ray is complete, perform the following:
a. Remove the x-ray cassette.
b. Return the x-ray cassette tray.
c. Place the warmer module in its normal operating position.

Universal Cylinder Bracket (Optional)

CAUTION:
When installing the cylinder, ensure that the universal cylinder bracket aligns the outlet of the
cylinder with the gas inlet on the back of the unit. Injury or equipment damage could occur.
The universal cylinder bracket can be adjusted to secure the cylinder in the proper position. It consists of
two metal parts: one inner fixed bracket attached to the lower post of the warmer and a sliding outer
bracket that mates with the cylinder. There is also a strap to secure the cylinder to the outer bracket.
1. Once the cylinder is properly installed, slide the outer bracket (A) away from the lower post until the
bracket meets the cylinder.
2. Tighten the nuts on the screw posts (B) located on the sides of the inner bracket until secure.
3. Wrap the cylinder strap (C) around the cylinder, close the latch, and pull on the strap until secure.