Dräger WBR82 Instructions For Use Manual page 54

Radiant warmer products
Hide thumbs Also See for WBR82:
Table of Contents

Advertisement

PROPRIETARY AND CONFIDENTIAL DRAFT 8 Jan 04
WARNING:
When transporting the infant in the cart and bassinet, place the mattress in the horizontal
position. Failure to do so could result in infant injury.
WARNING:
For patient safety and effective heating, properly dock and latch the cart to ensure correct
bassinet positioning.
WARNING:
Follow the product manufacturer's instructions. Failure to do so could result in infant injury,
personal injury, or equipment damage.
WARNING:
Only facility-authorized personnel should perform preventive maintenance or troubleshooting
on the Resuscitaire Radiant Warmer. Refer to the Resuscitaire Radiant Warmer Service Manual
for additional information. Preventive maintenance or troubleshooting performed by
unauthorized personnel could result in personal injury or equipment damage.
SHOCK HAZARD:
The potential for electrical shock exists with electrical equipment. Establish policies and
procedures to educate your staff on the risks associated with electrical equipment.
SHOCK HAZARD:
If the controller or other parts of the equipment fail to function properly, do not use the warmer. A
potential shock hazard exists within the warmer module. Refer service to qualified personnel.
Failure to do so could result in personal injury or equipment damage.
WARNING:
When the warmer module is swiveled and radiating heat, objects, such as monitors, located on
the optional monitor shelf may overheat or become hot to the touch. Personal or infant injury or
equipment damage could occur.
SHOCK HAZARD:
For servicing or transport, unplug the unit from its power source. Failure to do so could result in
personal injury or equipment damage.
SHOCK HAZARD:
Do not expose the unit to excessive moisture. Personal injury or equipment damage could
occur.
CAUTION:
When removing the equipment from the carton, use care not to scratch or otherwise damage
unprotected surfaces.
CAUTION:
Always use the specified filter between the suction canister and the suction tube coming from
the column. Failure to do so could result in equipment damage.