Dräger WBR82 Instructions For Use Manual page 73

Radiant warmer products
Hide thumbs Also See for WBR82:
Table of Contents

Advertisement

PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04
2. Install the wall bracket mounting bolts in the wall leaving 1/2" of the mounting bolts exposed.
NOTE:
The recommended wall bracket mounting bolts are 5/16" diameter. Four bolts are required, but are not
supplied with the equipment.
3. Mount the Resuscitaire Wall Mounted Radiant Warmer (WMRW82) on the bolts.
4. Tighten the bolts.
5. Install the wall mount cover plate onto the wall mount bracket using the four 6-32x1/4" screws.
6. Connect the AC power cord to the controller.