PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04
Error Number
Error #22
Error #23
Error #24
Error #25
7 - 4
Explanation
36° probe calibration error
>0.2°C
Ambient temperature too high
(>32°C)
There is an internal malfunc-
tion.
There is an internal malfunc-
tion.
Corrective Action
Adjust the temperature sensor of Skin Probe 1 and
Skin Probe 2.
Using an external thermometer, verify the ambient
temperature. If the temperature exceeds 32°C
(90°F), reduce the ambient temperature.
Refer the unit to service.
Refer the unit to service.