PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04

Cleaning

WARNING:
Follow the product manufacturer's instructions. Failure to do so could result in infant injury,
personal injury, or equipment damage.
WARNING:
If oxygen is in use, make sure that the oxygen supply to the equipment is turned off and
disconnected from the oxygen supply during cleaning and maintenance procedures. A fire and
explosion hazard exists when performing cleaning and maintenance procedures in an oxygen-
enriched environment.
SHOCK HAZARD:
For servicing or transport, unplug the unit from its power source. Failure to do so could result in
personal injury or equipment damage.
SHOCK HAZARD:
Do not expose the unit to excessive moisture. Personal injury or equipment damage could
occur.
CAUTION:
Do not use harsh cleansers, solvents, or detergents. Equipment damage could occur.
WARNING:
Cylinder heights may vary. When replacing a gas cylinder, ensure that the tank does not exceed
the maximum tank height specification. Also, prior to placing the cylinder, confirm that there is
sufficient clearance between the bottom of the tank and the floor when the VHA Resuscitaire
Radiant Warmer is in its lowest position prior to placement. After these checks have been
completed, raise the VHA to its maximum height and install/remove the cylinder. Failure to do
so could result in personal injury or equipment damage.