General Cleaning; Cleaning Agents; Steam Cleaning; Cleaning Difficult To Access Areas - Dräger WBR82 Instructions For Use Manual

Radiant warmer products
Hide thumbs Also See for WBR82:
Table of Contents

Advertisement

PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04

General Cleaning

At a minimum, thoroughly clean and disinfect the radiant warmer upon discharge of an infant. However,
depending upon individual facility policy, perform this as often as daily, if desired. Most effectively clean
by disassembling the unit and then grouping the parts and assemblies in categories according to the
method of cleaning required (see "Disassembly for Cleaning" on page 6-4).
We recommend that you clean the unit with detergent and warm water. Do not use excessive liquid or
harsh cleansers. Appropriate cleaning agents that will not damage the product or its accessories include
cold sterilants and quarternary ammonium compound disinfectants.
After cleaning, perform a complete operational checkout procedure (see "Operational Checkout
Procedure" on page 4-14).

Cleaning Agents

When the equipment is not in use and it is disassembled, use an intermediate-level (tuberculocidal)
detergent/disinfectant. Disinfect only when the equipment is not in use and disassembled (see
"Disassembly for Cleaning" on page 6-4). When using any cleaning agent, follow the manufacturer's
directions for use. Clean and disinfect the device using disinfectants and protocols approved by your
hospital. Dilute the disinfectant as specified on the manufacturer's label. Before cleaning, remove all solid
wastes and contaminants from the disassembled parts.

Steam Cleaning

Do not use any steam cleaning device on the unit. Excessive moisture can damage mechanisms in this
unit.

Cleaning Difficult to Access Areas

To remove difficult spots or stains, we recommend that you use standard household cleansers and a soft
bristle brush. To loosen heavy, dried-on soil, you may first need to saturate the spot.

Cleaning Painted Surfaces

Use disinfectant-detergent, such as cold sterilants and quarternary ammonium compound disinfectants to
clean all surfaces thoroughly. Make sure to clean all holes and indentations, and then dry with a clean
cloth or paper towel.

Cleaning Clear Plastic and Acrylic Surfaces

CAUTION:
Do not use alcohol, acetone, or any organic solvents for cleaning. Alcohol can cause crazing of
plastic and acrylic.
CAUTION:
Do not expose plastic and acrylic to direct radiation from germicidal lamps. Ultraviolet radiation
from these sources can cause cracking and crazing of clear plastic and acrylic.
1. Use disinfectant-detergent, such as cold sterilants and quarternary ammonium compound
disinfectants to clean all surfaces thoroughly. Make sure to clean all holes and indentations.
6 - 2