PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04
the alarm for 10 minutes, verify the infant's temperature, and press the Silence/Reset key. Failure to
acknowledge the Check Patient alarm by pressing the Silence/Reset key within the next 5 min causes
the heater to shut down, and an alarm occurs.
When operated in Baby Mode, the controller uses a skin temperature probe, connected between the
controller input and the infant, to automatically adjust the heater output of the warmer module to maintain
a selected set temperature.
The Apgar Timer displays the elapsed time and sounds an alert tone at 1, 5, and 10 minutes.
The Keypad Lock key, when pressed, enables or disables the Up/Down arrows, the Select Mode key,
and the >37°C key.
A procedural silence timer blocks out Low Baby Temp audible alarms for 5 minutes during routine
procedures.

Blender Module (Optional)

The blender module provides blended gas from 21% to 100% to the patient outlet on the resuscitation
module.

Resuscitation Module (Optional)

The resuscitation module contains pneumatic outlets necessary for infant resuscitation. Controls and
displays for the module are located above the bassinet on the warmer column.
Two variations of resuscitation modules exist: Resuscitation Module without Autobreath™ Infant
Resuscitator and Resuscitation Module with Autobreath™. The Resuscitation Module has an adjustable
airway pressure relief.
NOTE:
When pressure in the hose delivering gas to the patient falls below -0.4 kPa (-4 cm H
valve installed at the Patient Outlet connection opens. In the unlikely failure of the gas supply, the one-
way valve allows patient to inspire room air.
The resuscitation modules consist of the factory-installed components discussed on the following pages.