Mechanical Operation - Dräger Resuscitaire WBR82 Instructions For Use Manual

Radiant warmer product
Hide thumbs Also See for Resuscitaire WBR82:
Table of Contents

Advertisement

PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04
10. Check the Probe alarm:
a. Disconnect the skin probe from the skin probe connector. The Baby Temperature display goes
off, the Probe indicator flashes, and the alarm sounds.
b. Connect the skin probe to the skin probe connector.
11. Check the APGAR Timer:
a. Press the Start/Stop key. The APGAR Timer display comes on and begins counting up from 0 s.
b. Press the Start/Stop key. The APGAR Timer display stops counting.
c. Press the Reset key. The APGAR Timer display goes off.
12. Check the examination light:
a. Press the Examination Light key. The examination light comes on.
b. Press the Examination Light key. The examination light goes off.
13. Check the Power Failure alarm:
NOTE:
The unit must be connected to the AC line for at least 8 min before the power failure circuitry becomes
active.
a. Switch off the circuit breaker on the rear panel.
b. Switch on the On/Off switch on the front panel.
c. The Power Fail indicator comes on, and the alarm sounds.
d. Switch off the On/Off switch, and switch on the circuit breaker.
e. Disconnect the skin probe from the skin probe connector.

Mechanical Operation

1. Check the bassinet tilt control (Resuscitaire Birthing
Room Warmer only):
a. Support the rear lower edge of the bassinet with
your palm, and pull up on the lever located at the
bottom rear of the bassinet.
b. Place the bassinet in the 5 ° and then the 10 ° tilt
position.
c. Return the bassinet to the level position.
4 - 17

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Resuscitaire rw82Resuscitaire wmrw82

Table of Contents