10. Check the Probe alarm:
a. Disconnect the skin probe from the skin probe connector. The Baby Temperature display goes
off, the Probe indicator flashes, and the alarm sounds.
b. Connect the skin probe to the skin probe connector.
11. Check the APGAR Timer:
a. Press the Start/Stop key. The APGAR Timer display comes on and begins counting up from 0 s.
b. Press the Start/Stop key. The APGAR Timer display stops counting.
c. Press the Reset key. The APGAR Timer display goes off.
12. Check the examination light:
a. Press the Examination Light key. The examination light comes on.
b. Press the Examination Light key. The examination light goes off.
13. Check the Power Failure alarm:
NOTE:
The unit must be connected to the AC line for at least 8 min before the power failure circuitry becomes
active.
a. Switch off the circuit breaker on the rear panel.
b. Switch on the On/Off switch on the front panel.
c. The Power Fail indicator comes on, and the alarm sounds.
d. Switch off the On/Off switch, and switch on the circuit breaker.
e. Disconnect the skin probe from the skin probe connector.
Mechanical Operation
1. Check the bassinet tilt control (Resuscitaire Birthing
Room Warmer only):
a. Support the rear lower edge of the bassinet with
your palm, and pull up on the lever located at the
bottom rear of the bassinet.
b. Place the bassinet in the 5 ° and then the 10 ° tilt
position.
c. Return the bassinet to the level position.