1.29 Initializing and Calibrating the System (Initialization, reconfiguration, touch screen calibration), Setting the Measurement Mode
IM 04L51B01-01EN
Initializing a Specific Item
1
Press MENU.
The menu screen appears.
2
Tap the Browse tab and then Initialize Calibration.
The menu screen appears.
3
Tap Initialize > Settings/Internal data > Initialize all > Off.
4
To initialize the security settings, tap Security settings and then On. Otherwise, tap
^Off.
5
To initialize control settings, tap Control settings > On. Otherwise, tap Off.
6
To initialize the settings other than the security or control settings, tap Other than
Security settings and then On. Tap the items you want to initialize to set to On.
Otherwise, tap Off.
7
To initialize the data in the internal memory, tap Internal data and then On. Otherwise,
tap Off.
8
Tap Execute.
A confirmation screen is displayed.
9
Tap OK.
The settings are initialized. The message "Execution is complete." will appear.
Operation complete
Note
Before initializing, back up necessary setup parameters and other data.
Refer to page 1-238 in section 1.26.1, "Saving the Setting Parameters"
1
2
3
4
5
App
Index
1-251