Electrical Safety Inspection
The following electrical safety tests are recommended as part of a comprehensive preventive maintenance program. They
are a proven means of detecting abnormalities that, if undetected, could prove dangerous to either the patient or the
operator. Additional tests may be required according to local regulations.
All tests can be performed using commercially available safety analyzer test equipment. Please follow the instructions of the
analyzer manufacturer.
The consistent use of a safety analyzer as a routine step in closing a repair or upgrade is emphasized as a mandatory step if
an approved agency status is to be maintained. The safety analyzer also proves to be an excellent troubleshooting tool to
detect abnormalities of line voltage and grounding, as well as total current loads.
A.1 Power Cord Plug
Test Item
The power plug pins
The power
The plug body
plug
The strain relief
The power plug
The power cord
A.2 Device Enclosure and Accessories
A.2.1 Visual Inspection
Test Item
The enclosure and accessories
T1 Service Manual
Acceptance Criteria
No broken or bent pin. No discolored pins.
No physical damage to the plug body.
No physical damage to the strain relief. No plug warmth for device in use.
No loose connections.
No physical damage to the cord. No deterioration to the cord.
For devices with detachable power cords, inspect the connection at the device.
For devices with non-detachable power cords, inspect the strain relief at the device.
Acceptance Criteria
No physical damage to the enclosure and accessories.
No physical damage to meters, switches, connectors, etc.
No residue of fluid spillage (e.g., water, coffee, chemicals, etc.).
No loose or missing parts (e.g., knobs, dials, terminals, etc.).
A-1