6.6.11 Wired Network Related Problems

Symptoms
Unable to connect to the
wired network
The monitor is frequently
off line or disconnects from
the network.

6.6.12 Software Upgrade Problems

Symptoms
Possible Cause
Program
Incorrect network
upgrade
connection
fails
Wrong upgrade package
has been downloaded
Incorrect IP address setting
Main board defective
6-8
Possible Cause
Incorrect network cable
connection
Incorrect IP address setting
Network cable defective
Main board failed.
Incorrect network cable
connection
Incorrect IP address setting
Correction Action
1. Check the network connector on the patient monitor.
2. Make sure that the hub or switch is operating correctly. Check that correct
network cable is in use and has been connected correctly.
Upgrade package shall be .pkg files. Select the correct .pkg file for your application.
Configure a fixed IP address in range C as specified for the patient monitor. We
recommend not performing software upgrade whenT1 is connected to a network
having multiple PCs.
Replace the main board.
Correction Action
Check network cable connection. Network cable shall not be
longer than 50 m.
Check for IP address conflict. If yes, reconfigure the IP address.
Replace the network cable.
Replace the main board.
Check network cable connection. Network cable shall not be
longer than 50 m.
Check for IP address conflict. Reconfigure IP address.