7.4.3 Disassembling the Connector Assembly

Unscrew the two M3×6 screws indicated below. Slide the connector assembly slightly downward as shown below and then
lift the assembly to take it out from the top housing. Then disconnect the indicator cable from the connector assembly.
Note
When reassembling the assembly, make sure the silicon jacket is installed in place.
7.4.3.1 Removing Connector Board PCBA
Disconnect the power board cable from the connector board, unscrew the two M3×6 screws and take out the connector
board PCBA.
unscrew the two
M3×6 screws
T1 Service Manual
Disconnect the indicator cable
from the connector assembly
Unscrew the two
M3×6 screws
Disconnect the power
board cable from the
connector board
7-15