5.
Insert the antenna plugs into the Wi-Fi sockets (socket on the left for the cable on the left and socket on the right for the
cable on the right). Secure the antenna plug with tape.
6.
Route the cables and then reassemble the monitor.
7.
Turn on the monitor and then select [Main Menu]→[Maintenance>>]→[User Maintenance>>]→enter the required
password→[Network Setup>>], and set [Network Type] to [WLAN]. Test if the Wi-Fi function operates properly.

4.2 Software Upgrade

Software upgrades must be performed by Mindray, NA authorized service provider. Call Service Dispatch 1 800 288-2121 ext:
7875.