2.5.3 External Parameter Modules
T1 can be configured with the following external modules:
Sidestream CO
module
2
Microstream CO
2
T1 is connected with the external modules through the T1 Handle.
If T1 and an external module are both installed to the T1 handle, the external module is connected to the T1 handle converter
board through the T1 handle NIOS board gets power from the battery installed in T1. The T1 handle converter is connected to
T1 through the 54 pin connector. The T1 Handle can be installed to the T1 Docking station and be powered by the Docking
Station.
T1 Service Manual
module
Rack
NIOS
Board
Serial port
Rack
Conver
54pin
ter
Board
External Module
P
T1
l
Main Unit
u
g
2-11