7.3.3 Disassembling the Front Housing Assembly

1.
Unscrew the four PT3×10 screws that secure the front housing assembly and remove the front housing assembly from
the main frame. Make sure to disconnect the connectors for the Wi-Fi antenna, analog output cable, fan cable, DC-IN
cable, and analog signal cable. See the pictures below for details.
5GHz Wi-Fi module
has one antenna
only.
7-6
Unscrew the four PT3×10 screws
Cut the adhesive tape and
remove the Wi-Fi antenna
connector
Disconnect the fan cable
from this socket