Device Labeling; Scheduled Electrical Safety Inspection; Electrical Safety Inspection After Repair - Mindray T1 Service Manual

Patient monitor
Hide thumbs Also See for T1:
Table of Contents

Advertisement

A.2.2 Contextual Inspection
Test Item
The enclosure and accessories
A.3 Device Labeling
Check the labels provided by the manufacturer or the healthcare facility are present and legible.
Main unit label
Integrated warning labels
A.4 Scheduled Electrical Safety Inspection
For scheduled electrical safety inspection, perform all the test items listed in A.6 ELECTRICAL SAFETY INSPECTION TEST
A.5 Electrical Safety Inspection after Repair
The following table specifies test items to be performed after the equipment is repaired. Refer to A.6 ELECTRICAL SAFETY
INSPECTION TEST for the description of the test items.
Repair with main unit not disassembled
Repair with
When neither power supply PCBA nor patient
main unit
electrically-connected PCBA is repaired or
replaced
disassembled
When power supply PCBA is repaired or
replaced
When patient electrically-connected PCBA is
repaired or replaced
When both power supply PCBA and patient
electrically- connected PCBA are repaired or
replaced
A-2
Acceptance Criteria
No unusual noises (e.g., a rattle inside the case).
No unusual smells (e.g., burning or smoky smells, particularly from ventilation holes).
No taped notes that may suggest device deficiencies or operator concerns.
Test items: 1, 2, 3
Test items: 1, 2, 3, 4
Test items: 1, 2, 3, 4, 5
Test items: 1, 2, 3, 4, 6, 7, 8
Test items: 1, 2, 3, 4, 5, 6, 7, 8
.