1.1.1 Warnings

WARNING
All installation operations, expansions, changes, modifications and repairs of this product are conducted by
authorized personnel.
Disconnect the patient monitor from external power source and remove the battery before disassembling the
equipment.
When you disassemble/reassemble a parameter module, a patient leakage current test must be performed
before it is used again for monitoring.
The equipment must be connected to a properly installed power outlet with protective earth contacts only. If the
installation does not provide for a protective earth conductor, disconnect it from the power line and operate it
on battery power, if possible.
Dispose of the package material, observing the applicable waste control regulations and keeping it out of
children's reach.

1.1.2 Cautions

CAUTION
Make sure that no electromagnetic radiation interferes with the performance of the equipment when preparing
to carry out performance tests. Mobile phone, X-ray equipment or MRI devices are a possible source of
interference as they may emit higher levels of electromagnetic radiation.
Before connecting the equipment to the power line, check that the voltage and frequency ratings of the power
line are the same as those indicated on the equipment's label or in this manual.
Protect the equipment from damage caused by drop, impact, strong vibration or other mechanical force during
servicing.

1.1.3 Notes

NOTE
Refer to Operation Manual for detailed operation and other information.

1.2 Equipment Symbols

See the T1 Operator's Manual for information about the symbols used on this product and its packaging.
1-2
T1 Service Manual