Temp Test; Ibp Performance Test; Co Tests; Co 2 Leakage Test - Mindray T1 Service Manual

Patient monitor
Hide thumbs Also See for T1:
Table of Contents

Advertisement

5.3.7 Temp Test

Tool required:
Resistance box (with accuracy above 0.1Ω)
Follow this procedure to perform the test:
1.
Connect the two pins of any Temp connector of a module to the two ends of the resistance box using 2 wires.
2.
Set the resistance box to 1354.9Ω (corresponding temperature is 37ºC).
3.
Verify that the displayed value is within 37 ± 0.1ºC. If the temperature is not within 37 ± 0.1ºC, contact Mindray Technical
Support.
4.
Repeat steps 1 to 3 and verify another temperature channel.

5.3.8 IBP Performance Test

Tools required:
Medsim300B patient simulator, or MPS450, or equivalent equipment
IBP adapter cable for test (P/N 009-002199-00 for Medsim 300B, P/N 009-002198-00, for MPS450)
Follow this procedure to perform the test:
1.
Connect the patient simulator to the monitor's IBP connector.
2.
Verify the patient simulator output to the IBP channel is zero.
3.
Select IBP Zero in the IBP setup menu to make a zero calibration.
4.
Configure the patient simulator as P (static) = 200 mmHg.
5.
The displayed value should be within 200 ± 4 mmHg. If the error is beyond ±4 mmHg, return the IBP module to the
factory for repair.
6.
Set the patient simulator output to 120/80 mmHg ART signal and 120/0 mmHg LV signal to the IBP channel and check
that the IBP wave is displayed correctly.
7.
Repeat the steps above for all the IBP channels.
5.3.9 CO
Tests
2
See 5.2.2 CO2 Leakage Test, 5.2.3 CO2 Accuracy Test, and 5.2.4 CO2 Calibration.
T1 Service Manual
5-9